Welcome to the RCT Console Web Site!

RCT Console is a powerful Device Type Manager (DTM) for Badger Meter Inc. Coriolis Mass Flow Meters. The program provides instrument configuration, multi-parameter data logging/graphing, asset optimization through remote/local maintenance, historian functions, and specialized functional tools (such as batching, backup and restore, etc.). The program implements instrument access by BMI Protocol via RS485 and RS232 connection. Bluetooth and USB serial ports are also supported.

Friday, January 4, 2019
Download the latest RCT Console software for your RCT1000 transmitter. (9.19.10121.1545)
Friday, January 4, 2019
Visit the latest help file on-line on the RCT Console Help System site.
What's new in the latest release?
RCT Console newsletter

What's new in latest release of RCT Console?

Version 9.17.11214.1730 - December 14, 2017.

New features:

  • Implement 10-point sensor linearization manager tool. (Automated Meter Calibrator)
  • Implement New Calibration from Template File function. With this function an existing calibration file can be open but only the calibration plan will only kept. Using this function you can re-do any existing calibrations without creating manually a new CALX file with a new calibration plan. (Meter Calibrator, Automated Meter Calibrator)
  • Implement Automated Meter Calibrator feature for AFS MassMeter transmitter. This makes possible to automated calibration with a connected scale, current controlled pump, and connected emptying valve. All hardware elements are controlled by the calibrated transmitters digital outputs.
  • Show current transmitter name in "[ID], [TransmitterName] - Parameter Access Toolbar" format (PArameter Access Toolbar)
  • Add a new feature which adds a new variable (to the current parameter) based on the already existing variable of the parameter. (Configuration Editor of Transmitters.)
  • Read Transmitter button - in Parameter Access Toolbar of the Transmitter Explorer Ribbon tab - got a dropdown list with the most frequently used parameters (maximum 10). The recent parameters list is "dynamic", the most frequently used parameter will always be the first element in the list.

Bug fixes:

  • Resolves an issue with creating batch setpoint in wrong unit when creating new calibration. (Meter Calibrator)
  • Resolves an issue with losing the focus from parameter selector drop-down list box after reading the parameter value. With this fix the parameter selector drop-down list box gets back the focus after reading the parameter value from the transmitter and you can type in the next parameter ID number without needing to click into the list box again. (Parameter Access Toolbar)
  • Resolves an issue with causing run-time error if count of last used parameters is greater than 11. (Parameter Access Toolbar)
  • Resolves an issue with faulty executing scroll bar zooming in Graph. (Graph/Data Logger)
  • Resolves an issue with showing faulty standard deviation values in Graph. (Graph/Data Logger)
  • Resolves an issue with connecting to Sartorius scale caused run-time error. (Meter Calibrator)
  • Resolves an issues with causing "?" characters in localized text on the LCD display of transmitter (RCTX-1000)
  • Resolves an issue with not showing always parameter values in Transmitter Browser if more than one, but same type transmitter are added to the Browser. (Transmitter Browser)
  • Correct the LCD character filtering for ROM code 0E in Localization function (RCTX/N-1000 transmitters)
  • Resolves an issue with causing run-time error adding a new parameter in Configuration Editor of Transmitters.

Maintenance

  • RCT Console is built with the latest (2017V3) ComponentOne control set.

9.16.10207.1530 - February 7, 2017

RCT Console
  • Resolves an issue with causing run-time error when record type parameters are added to restore parameters list.

9.16.11219.1730 - December 19, 2016

RCT Console
  • Contains localized (text and screen shots) French, German and Spanish help files.
  • Uses the latest 2016V3 C1 control set.

9.16.11109.1145 - November 9, 2016

This version contains bug fixes.

RCT Console
  • Resolves and issue with showing could not save RCT Console Historian file error when closing RCT Console.
  • Clicking on RCT Console Historian button in the status bar opens the RCT Console Historian.

9.16.10830.1400 - August 30, 2016

This version contains bug fixes and several new features. Please note that this is a public pre-release candidate version so some languages might not be fully supported.

RCT Console
  • Localization of the Application added (English, French, German, Spanish, Hungarian)
  • Localization of Transmitter Configuration editor added (English, French, German, Spanish, Hungarian)
  • Uses the latest 2016V2 C1 control set
Transmitter Tester
  • Supports testing RCTN and RCTX type transmitters
  • “Quick and dirty” type test mode added.
  • Log file created with overall test results.
  • Default values of several test revised.
BMI RCT1000 Transmitter
  • Localize transmitter function added. (English, French, German, Spanish, Hungarian)
DualGraph
  • Allows changing units on X axis (seconds, minutes, hours, days)
  • Allows plotting parameters on Y1(left) and/or Y2 (right) axes
Meter Calibrator
  • Allows using constant current output mode for PID Controller with adjustable pre/post calibration waiting time.

9.15.11013.1330 - October 13, 2015

RCT Console
  • This edition does not contain any change in RCT Console.
Meter Calibrator
  • Resolves an issue with starting Meter Calibrator caused run-time error in release mode compiled WSD9AfsMassmeter.dll.

9.15.10807.1820 - August 7, 2015

RCT Console
  • The application is digitally signed with Badger Meter's Code Signing Certificate.
  • The Transmitter Manager shows state of the COM port(s) assigned to transmitters (Icon and text).
  • If no communication port found in the computer during startup a new COM Port Warning Dialog is shown. The warning dialog allows the user to turn on Communication Simulation, or run the program without COM Port. The Communication Simulation is not enabled automatically any more.
  • New Options menu icons
  • Uses the latest 2015 V2 C1 control set
  • Resolves an issue with C1 Reporting controls

9.15.10419.1845 - April 19, 2015

RCT Console
  • Both the upper (Primary) and the lower (Secondary) charts have a second Y axis. Any parameter can be displayed either on Y1 or Y2 axis each chart. Both Y axes are individually auto scaled by default.
  • Scrollbar Zooming function added. When this function is turned on (in Graph Toolbar) then RCT Console shows a vertical and a horizontal scrollbar nearby the Y and X axes and the chart can be "scrolled" according to the set zoom scale: 1:100 (default for X axis), 1:50, 1:20, 1:10, 1:5, 1:2 and 1:1 (default for Y1 and Y2 axes)
  • Show and Reset Password buttons added to RCT Console Password dialog.
  • Options form and menu button use the popular "Hamburger" icon.
  • This version is recompiled with the latest control set (2015V1)
  • The installer, the executable and library files are stamped with Badger Meter's code signing certificate.
  • Resolves an issue with the App was closed even if the user cancelled saving the changed graph.
  • Resolves an issue with saving and restoring dual graph state (I.e. single or dual graph was used)
Meter Calibrator
  • Resolves an issue with scrambled the Error graph options during saving the CALX file.
  • Resolves an issue with not showing the error trumpet curve up to the maximum flow rate of the sensor only to the maximum flow rate used during the calibration.

9.14.11002.1800 - October 2, 2014

RCT Console
  • Resolves an issue with not showing the simulated value of parameters in Graph if communication simulation is enabled.
Meter Calibrator
  • Resolves an issue with not always saving results and calibration information if the Density Calibrator Wizard is closed with Finish button.

9.14.10902.1015 - September 2, 2014

Meter Calibrator
  • Pass/Fail icon shown for each measurement points in Meter Calibrator form and printed report.
  • Meter factors in Edit Calibration Information are write protected by default.
  • Upload Constants button added to upload meter factors from current transmitter.
  • Sensor Size Editor form added to browse, edit or modify default meter factors.
  • Sensor error and Zero Stability error limits added to Sensor Size Editor.
Bug fixes
  • Resolves an issue with not saving checked state of AutoUpdate check box in Options
  • Resolves an issue with having hard coded form title (not localized) of Update Dialog
  • Resolves an issue with several parameters could not be backed up and/or restored

9.14.10527.1300 - May 27, 2014

RCT Console
  • Spelling of 177, Baume fixed
Meter Calibrator
  • Buttons placed into Quick Access Toolbar of Ribbon are saved and restored next time.
  • New configuration file (rcsSizeTable.xml) containing the latest sensor size table was added to Configuration folder.
  • Resolves an issue with not saving changes if Edit Calibration Information dialog was closed by pressing enter key on keyboard.
  • Resolves an issue with not saving sensor size after closing Edit Calibration Information dialog.
  • Resolves an issue with printing Flow Factor = 0.0 in calibration report

9.14.10509.1700 - May 9, 2014

RCT Console
  • Resolves an issue causing run-time error when HealthTrack dialog is closed during collecting data from transmitter
  • Logging read-protected parameters from RCT-1000 type transmitter is supported in HealthTrack

9.14.10506.1000 - May 6, 2014

RCT Console
  • New red style RCT Console, RTC-1000 and Transmitter Simulator icons added.
  • Label of parameter 5 and parameter 202 fixed.
Meter Calibrator
  • Resolves an issue with not saving the sensor size parameter value into the CALX file.
  • If you open a CALX file which does not contain the sensor size info then Meter Calibrator "tries to find out" the sensor size from the other available sensor data.

9.14.10403.1100 - April 3, 2014

RCT Console
  • Help file updated with new RCT Console document folder.
  • Built with ComponentOne 2014V1 controls set.
  • When changing unit of a record type variable then unit of the parent parameter is also changed.
  • Folder name changed  from "..\Documents\ScrewDRIVER9"  to "..\Documents\Badger Meter\RCT Console".
  • Changing unit system (Parameter 296) does not require password.
  • 299, APP Program Control parameter added which dynamically controls which APP Programs are executed.
  • 300, Menu scroll direction parameter to added which dynamically controls the scrolling direction of menus in transmitter.
  • Description of parameters shown in the status bar of Transmitter Browser.
Meter Calibrator
  • Volumetric (tank) and/or Master Turbine Meter based calibrations supported.
  • When crating new calibration the parameter 16, Sensor Size can be set which automatically updates the flow factor, flow max/mmin, and density constants and transmitter serial number.
  • When crating new calibration the last used laboratory and calibrators names are saved.
  • Double clicking on batch size cell in calibration sheet starts in-line editor instead of showing the whole measurement point.
  • When entering measured volume the density of water is automatically calculated using method developed by International Committee for Weights and Measures (CIPM) 2001.
  • Density of water are different in every calibration lab. A density offset can be added and stored for any calibration lab.
  • Batch Type parameter is changed according to the current calibration type when starting the calibration.
  • The necessary batch APP Program is selected when starting a calibration.
  • Resolves an issue with not saving pre and over sampling maximum values into the CALX file.
Density Calibrator
  • Density is not shown in the data sheet during density calibration in air nor in water.
  • When exporting density calibration data the density alpha and beta constants are also exported

9.14.10217.1100 - February 17, 2014

RCT Console
  • RCT Console web site created
  • RCT Console uses the new RTC Console web site to check for updates.
  • Jump to Online Help web site link is added to Help Menu
  • Jump to RCT Console web site link is added to Help menu
Meter Calibrator
  • Resolves and issue with starting Density Calibrator causes run time error if DensityDW value > 100 in any unit.

9.14.10109.1430 - January 9, 2014

RCT Console
  • K-Flow (ABB, Afs 1210) transmitters are supported now
  • Resolves an issue with showing scroll bars in WSD9RegistrationDialog.
  • Resolves an issue with decrementing the decimal digits if the number of decimals were already 0.
  • Resolves an issue with showing "Aborted by user" error message when pausing HealthTrack.
  • Resolves an issue with disabling "Show splash screen during startup" does not disable the splash screen next time
  • Resolves an issue with showing "Build at" instead of "Build on" in About RCT Console and the splash screen
BMI Simulator
  • Operators can generate unique serial number for each new transmitter automatically in Transmitter Tester APP
Meter Calibrator
  • Resolves an issue with not printing calibration file name in report header
  • Resolves an issue with printing Meter Info data left aligned and unit into a new column
  • Resolves an issue with printing the default nominal Flow Factor value not the calibrated one
  • Resolves an issue with showing download Meter constants dialog after closing Edit Calibration Information dialog even though just the calibration information was edited.
  • Resolves an issue with having value.Decimals = -1 causes the value is printed as "F-1" in report
  • Resolves an issue with showing not executed/empty measurement points in report
  • Resolves an issue with disabling the scale if auto-connecting or reading the scale failed
  • Resolves an issue with updating the scale value during zero calibration in Meter Calibrator

9.13.11211.1300 - December 11, 2013

  • BMI RCT1000 Transmitter - Meter Calibrator
    • Resolves an issue with not showing measurement points with FlowFactorCalibration = true in error graph
    • Resolves an issue with not showing the delete button in Meter Calibrator's recent files if the file name was too long
    • Resolves an issue with showing duplicated group IDs in Meter Calibrator's measurement points list
    • Resolves an issue with showing empty calibration report if the calibration report definition file not found
    • Edit calibration point dialog contains a check box to flag the edited calibration point as Flow Factor Calibration point
    • Edit Calibration Information dialog - Error graph tab contains a check box to show or hide the measurement points marked as Flow Factor Calibration points in error graph

9.13.11210.1200 - December 10, 2013

  • RCT Console bug fixes
    • Resolves an issue with showing "Build at" instead of "Build on" in About RCT Console and the splash screen
    • Resolves an issue with disabling "Show splash screen during startup" does not disable the splash screen next time
    • Resolves an issue with showing "Aborted by user" error message when pausing HealthTrack.
  • BMI RCT1000 Transmitter
    • Resolves an issue with updating the scale value during zero calibration in Meter Calibrator
    • Resolves an issue with disabling the scale if auto-connecting or reading the scale failed in Meter Calibrator

9.13.11210.1200 - December 10, 2013

  • RCT Console bug fixes
    • Resolves an issue with showing "Build at" instead of "Build on" in About RCT Console and the splash screen
    • Resolves an issue with disabling "Show splash screen during startup" does not disable the splash screen next time
    • Resolves an issue with showing "Aborted by user" error message when pausing HealthTrack.
  • BMI RCT1000 Transmitter
    • Resolves an issue with updating the scale value during zero calibration in Meter Calibrator
    • Resolves an issue with disabling the scale if auto-connecting or reading the scale failed in Meter Calibrator

9.13.11126.1130 - November 26, 2013

  • RCT Console
    • Resolves an issue with ModBus protocol receiving shorter string than expected.
    • Controls updated to 2013V3 version
    • Resolves an issue with inconsistently using visual styles of ComponentOne controls
    • Resolves an issue with "Save to Excel" button showing wrong tool tip text in WSD9Simualtor.WSD9FormTxmTester form
    • Resolves an issue with showing Symbol "SiF" instead of "SlugFlo" of 119, Slug in Flow parameter of BMI Transmitter

9.13.11119.1130 - November 19, 2013

  • RCT Console
    • Issue fixed: Deleting the last transmitter form Transmitter Manager caused run-time error.
    • Issue fixed: Transmitters' list can only be modified if all the basic and/or special functions are closed.
    • Issue fixed: Title of Transmitter not found dialog.

9.13.11011.1000 - October 11, 2013

  • RCT Console
    • Issue fixed: Aborting AutoUpdate dialog might cause run-time error.

9.13.10907.1100 - September 7, 2013

  • RCT Console
    • New feature added: Item List Setup Editor automatically selects parameters in the source parameters list. This feature works only if user enters a comma delimited list into the search box. (i.e. "33,59,60,65")
    • New feature added: Find parameter function searches in symbol of the parameters.
    • New feature added: Find parameter function has a Clear button
    • Issue fixed: Pressing F1 in main window opened the Help but with "Page not found" error.
    • Issue fixed: RCT Console password could accidentally disappear.
    • Issue fixed: The default file name of Graph was ".dlgx" not "Untitled GraphXXX.dlgx"
    • Issue fixed: Layout of Graph/Data Logger dialogs saved corrupt data accidentally and caused blank forms.
  • BMI RCT1000 Transmitter
    • New transmitter icons added
    • Issue fixed: When Meter Calibrator was started all the control buttons were enabled.
    • Issue fixed: Meter Calibrator Data Logger function did not save the axes color information into the CALX file and caused blank charts.
  • BMI Simulator
    • New transmitter icons added

9.13.10903.1230 - September 3, 2013

  • RCT Console
    • New feature added: When closing the application the Make Flash Backup Dialog is shown if any flash backup protected item changed.
    • Issue fixed: HealthTrack failed to generate automatic file name if the transmitter serial number contained '/' or '.' characters or extra spaces.
    • Issue fixed: COM History did not show the Item View right after changing RFI Protocol name to BMI Protocol.
  • BMI RCT1000 Transmitter
    • New transmitter icons added
    • Issue fixed: 137, Alarm-F low limit had wrong unit
    • New feature added: User defined items for the Home screen implemented (#700-#707)
    • New feature added: Parameter #295, Count of changed parameters implemented
  • All transmitter servers
    • New feature added: IsConfigurationChanged() and MakeFlashBackup() functions implemented.

V9.13.10822.1000 - August 22, 2013

  • RCT Console
    • New feature added: RCT Console Password. Some special functions and Transmitter Configurator are password protected.
  • BMI RCT1000 Transmitter
    • RFI protocol renamed to BMI Protocol
    • RFI MassMeter transmitter renamed to BMI RCT1000 Transmitter
    • MassTrack function renamed to HealthTrack
    • MassTrack folder renamed to HealthTrack. All the new HealthTrack records will be save into the new HealthTrack folder.
    • Help file updated. It uses BMI Protocol, BMI RCT1000 Transmitter, HealthTrack and the screen shot of new HMI menu system driver Remote Display

V9.13.10812.1000 - May 12, 2013

  • RCT Console
    • New feature added: Parameter Symbol LCD Display in Parameter Access Toolbar.
      Note: The Parameter Symbol LCD Display is shown only if the symbol text is defined in Transmitter Configuration file.
  • Transmitter Configurator
    • New feature added: Parameter Symbol editor box in Variable View. (To add, edit delete parameter symbols)
  • MassMeter transmitter
    • New feature added: HMI menu system compatible Remote Display and keyboard handling.
    • New feature added: PID controller based flow calibration points. User can set the desired flow rate and enable the PID Controller for each calibration points and the flow rate will be set automatically.
      Note: A current controlled pump is necessary for this feature.
    • Issue fixed: Calibration report printed wrong column names and data

V9.13.10517.1000 - May 17, 2013

  • RCT Console
    • Issue fixed: Run-time error occurred if user started data logging with partially specified file name in Graph Wizard.
    • Issue fixed: Localization of Parameter List Editor fixed
  • MassMeter transmitter
    • Issue fixed: Run-time error occurred if user started Flow Rate Gauge in Meter Calibrator APP.

V9.13.10412.1035 - April 12, 2013

  • RCT Console
    • Recompiled with WSD9Simulator.dll
  • Transmitter Tester
    • Issue fixed: Old version of C1Excel.dll was used

V9.13.10411.1040 - April 11, 2013

  • RCT Console
    • Localization, images and icons for Badger Meter, Inc. added
    • ComponentOne controls updated to 2013 V1
    • Issue fixed: Units mismatch in Transmitter Browser.
    • Issue fixed: Transmitter Configuration Editor locked in XML view.
  • RCT1000 Transmitter
    • New feature: Support for Ticket Printer control parameters
  • Transmitter Tester
    • Issue fixed: Transmitter and Simulator serial number mismatch

V9.13.10222.2000 - February 22, 2013

  • RCT Console
    • Issue fixed: After closing RCT Console Historian a run-time error occurred when user modified a Historian enabled parameter.
    • Issue fixed: Output minimum text was shown instead of Output maximum for analog output setup parameters.
    • Issue fixed: Record type parameters could not be edited in Transmitter Browser.
    • Issue fixed: Closing Transmitter Browser while a parameter is edited causes run-time error.
  • Transmitter Tester
    • Issue fixed: Data logger of transmitter tester stopped working if there was no reply received from the Simulator for a write command.

V9.13.10219.1000 - February 19, 2013

  • RCT Console
    • Transmitters: Using unsigned long data type parameters are supported
    • Installation: Windows 8 is supported.
    • Installation: Evaluation version dialog removed from InstallShield wizard

V9.13.10208.1220

  • RCT Console
    • RCT Console Historian APP added. RCT Console Historian saves any changes made to any historian enabled parameter from any transmitter to a database. Using the new Historian dialog you can open the database and sort, filter the changes made to your transmitters Historian protected parameters. Historian database can be exported to Microsoft Excel format.
    • Default list of historian enabled parameters for MassMeter and Simulator updated.
    • Issue fixed: Clicking on Transmitter Type button in task bar did not start the Transmitter Editor.
    • Issue fixed: MassTrack APP used the serial number of transmitter instead of the sensor when generated the MassTrack file name.
  • Transmitter Simulator
    • Fixed issue: Run time occurred when Transmitter Tester form was closed and the Restore parameters button was clicked.
    • Fixed issued: User could enable/disable any RTD tests regardless the RTD setup was Write protected.
  • All Transmitters
    • RCT Console Historian support added to all the transmitter plug-in modules.
  • MassMeter transmitter
    • Compatible with the latest firmware (M1.13.02071050)
    • Sensor(6) and transmitter(5) serial number parameters support the 12 character long serial numbers
    • Sensor Type(4) parameter is read only
    • Sensor Size(16) parameter supports the latest RFI sensor sizes

V9.13.10118.1445

  • RCT Console
    • Better and safer initialization methods added to data logger function of Transmitter tester app.
    • Check for Updates function shows the version of the available update and the version of the user's software
    • Version number comparing issue fixed in Update Library

V9.13.10117.1315

  • RCT Console
    • Recompiled version with RFI help engine

V9.13.10115.1030

  • RCT Console
    • WSD9AfsMassMeter transmitter displays "RFI MassMeter" transmitter name and "RFI Protocol" protocol name if compiled for RFI.
    • WSD9Simulator transmitter displays "RFI Simulator" transmitter name and "RFI Protocol" protocol name if compiled for RFI.
    • Parameter 200, Batch preset 6 was added to WSD9AfsMassMeter transmitter
    • Default error limit of Analog Input test was increased to 100mV in Transmitter Tester app.

V9.12.11212.1300

  • RCT Console
    • This release uses the latest 2012 V3 components
  • Meter Calibrator
    • Fixed issue: OmegaC was always saved as 1.0 regardless of the calculated OmegaC

V9.12.11204.1200

  • RCT Console
    • Revised find parameters function
  • Meter Calibrator
    • Digital IO test added for EXD transmitters
    • Start/finish date stamp and elapsed time are showed at the end of the automated test
    • Test report folder can be specified in general tab. Meter Calibrator stores the new test/excel export files in the specified location.

V9.12.11203.1430

  • Meter Calibrator
    • Recent documents handling added to the File menu of ribbon
    • Show write protected check box on the top of RTD setup form like on other setup forms
    • Fixed issue: Not all test showed the "Test was successfully aborted by user" dialog
  • Sensor Simulator
    • Fixed issue: If RTD slope in transmitter was zero then run-time error occurred when the RTD setup was shown
    • Fixed issue: Excel report if made by automated test did not contain the header information

V9.12.11202.2030

  • RCT Console
    • Unit of Batch counter, amout/time could not be changed from Parameter Access toolbar.
  • Meter Calibrator
    • Average flow rate and percentage in a flow test pass are calculated using histogram now. (Maximum of valid samples +/-%5 range)
  • Configuration editor
    • Default Metric and English unit setting support added
  • Sensor Simulator
    • Automated Transmitter Test implemented
    • EXD/NEMA4 switch added
    • Fixed issue: After a completed an Digital IO test the Digital Input #4 or #2 left ON depending to the NEMA4/EXD switch state
    • Fixed issue: Progress bar showed faulty batch counter percentage if the unit of batching was [lb].
    • Fixed issue: Run-time error occurred if user closed the tester form and he clicked on Backup button.
    • Fixed issue: Opening a test file which had only Analog#1 test caused run time error
    • Fixed issue: Transmitter Tester showed wild values during RTD test if the Temperature unit was °F
    • Fixed issue: Temperature error limit could not be modified if the Temperature unit was °F
    • Temperature Error dimension added. The temperature error in °F = 1.8 * error in °C (without the +32 °F offset)

V9.12.11128.1200

  • RCT Console
    • Fixed issue which caused showing "Missing Transmitter Configuration files" error dialog after installing RCT Console on a new PC (Not upgrading or updating an existing version).
    • Fixed issue which caused the user had to find, download and install Microsoft .NET 3.5 SP1 manually. This is done by the installer automatically.

V9.12.11127.1000

  • RCT Console
    • Application menu (top left side blue arrow) of ribbon control is renamed to "Tools" menu
    • Customized Quick Access Toolbar items in ribbon control are saved and restored in the next session
  • Sensor Simulator
    • Export to Microsoft Excel function added
    • Issue solved: Drive level test left the drive level at 100% after the test was completed.
    • Issue solved: Phase test setup showed "F-1" instead of the Frequency set in Simulator
  • AFS Mass Flow Meter transmitter (Meter Calibrator)
    • Application menu (top left side blue arrow) of ribbon control is renamed to "File" menu
    • Both the absolute and zero stability error limits can be modified in the new Error Graph tab of Edit Calibration Information dialog.
    • The error graph of meter calibrator can show the error limits in absolute flow rate units or in percentage of maximum flow.

V9.12.11027.1000

  • RCT Console
    • Communication status LED was red during remote display (Command 'D') or data logger memory upload communications (Command 'M')
    • Remote Display shows the status characters of zero calibration('z') and Batch Controller ('b' or 'B')
    • Remote Display shows the error level character of transmitters
    • Configuration files are checked during every startup
  • Sensor Simulator
    • Default values of all tests revised
    • Digital I/O test added
    • Frequency output test added
    • Analog Inputs test added
    • All tests' setup are write protected by default.
    • Backup/Restore constant to/from flash dialog added

V9.12.10927.1000

  • RCT Console
    • Documents\ScreDRIVER9\Updates folder added to manage the latest updates
  • Update Service
    • Improvements of safer updates
    • More detailed error dialogs
    • Updates are downloaded to the user's Documents\ScreDRIVER9\Updates folder.
  • Sensor Simulator
    • Analog outputs test added

V9.12.10924.1215

  • RCT Console
    • Transmitter error LED is always shown if the protocol is AFS Massmeter. If the transmitter does not support the Error Record parameter then clicking on error LED does nothing
    • Transmitter Manager: Baud rate is left aligned
    • Transmitter Manager: double click on transmitter name starts the in-line transmitter name editor
    • Transmitter Manager columns are auto sized after adding/deleting/renaming/editing transmitters
    • Add new transmitter dialog shows the selected transmitters type name as transmitter name by default. If user changes the type of the transmitter then the name of the transmitter is automatically changes.
  • AFS Mass Flow Meter transmitter
    • Changing any unit in Meter calibrator ensures that the whole sheet will use the same unit automatically.
    • If the unit of Batch size is changed then unit of Batch Counter and Corrected Scale will also be changed automatically. This works also if you change the unit of Batch Counter or Corrected Scale.
    • Error calculated as 100 * (batch counter - corrected scale) / corrected scale
    • New flow factor is calculated as FFold * AverageOf(corrected mass) / AverageOf(batch counter))
    • During opening existing CALX files the error is checked and fixed if it is necessary.
    • Measurement points belong to Flow factor calibration will not be shown on error graph
  • Sensor Simulator
    • Drive Response test added.